Keithley 2304A

Nguồn DC keithley 2304A, một ngõ ra tùy chỉnh 100W

Thông tin sản phẩm:

The Model 2303/2304A Power Supplies provide both voltage control and power consumption monitoring for automated testing of portable, battery-operated devices. These DC power supply products are optimized for testing battery-operated, wireless communication devices such as cellular phones that undergo substantial load changes for very short time intervals. These power supplies exhibit outstanding voltage stability during pulse load changes and can simultaneously measure load currents, even if they are short pulses. In addition, this family of power supplies can sink current and, thus, take on the characteristics of a discharged, rechargeable battery for testing chargers and charger-control circuitry.

 

Thông tin bổ sung

Hãng SX

Keithley

GENERAL SPECIFICATIONS
ISOLATION (low-earth) 22VDC max
PROGRAMMING IEEE-488.2 (SCPI)
USER-DEFINABLE POWER-UP STATES 5
REAR PANEL CONNECTOR 8-position quick disconnect terminal block for output (4), sense (2), and DVM (2)
TEMPERATURE
COEFFICIENT
(outside 23°C ±5°C)
Derate accuracy specification by (0.1 × specification)/°C
OPERATING TEMPERATURE 0° to 35°C (Full power).
0° to 50°C (Derate to 70%)
STORAGE TEMPERATURE –20° to 70°C
HUMIDITY <80% @ 35°C non-condensing
POWER CONSUMPTION 150VA max
REMOTE DISPLAY/KEYPAD OPTION Disables standard front panel
DIMENSIONS 89mm high × 213mm wide × 360mm deep (3 1 / 2 in × 8 1 / 2 in × 14 3 / 16 in)
NET WEIGHT 3.2kg (7.1 lbs)
SHIPPING WEIGHT 5.4kg (12 lbs)
INPUT POWER 100–120VAC/220–240VAC, 50 or 60Hz (auto detected at power-up)
WARRANTY One year parts and labor on materials and workmanship
EMC Conforms with European Union Directive 89/336/EEC EN 55011, EN 50082-1, EN 61000-3-2 and 61000-3-3, FCC part 15 class B
SAFETY Conforms with European Union Directive 73/23/EEC EN 61010-1, UL 3111-1
AC LINE LEAKAGE CURRENT 450µA @ 110VAC, typ.; 600µA @ 220VAC, typ
RELAY CONTROL JACK 1-channel, sink 150mA max., 15V max. 5V output, 100mA max., also available on jack. Accepts 0.173 in Bantam-type plug (CS-1003-1)
ACCESSORIES SUPPLIED User manual, service manual, output connector mating terminal (part no. CS-846)
ACCESSORIES AVAILABLE Model 2304-DISP Remote Display/Keypad (4.6 in × 2.7 in × 1.5 in). Includes 2.7m (9 ft) cable and rack mount kit
Optional Version Model 2303B 2303 with blank front panel (only AC power indicator LED)

 

DC VOLTAGE OUTPUT (2 Years, 23°C ± 5°C)
OUTPUT VOLTAGE 0 to +15VDC
OUTPUT ACCURACY ±(0.05% + 10mV)
PROGRAMMING RESOLUTION 5mV
READBACK ACCURACY ±(0.05% + 3mV)
READBACK RESOLUTION 1mV
OUTPUT VOLTAGE SETTLING TIME 5ms to within stated accuracy
LOAD REGULATION 0.01% + 2mV
LINE REGULATION 0.5mV
STABILITY 0.01% + 0.5mV
TRANSIENT RESPONSE TO 1000% LOAD CHANGE Transient Recovery Time <40µs to within 100mV of previous level.
<80µs to within 20mV of previous level
Transient Voltage Drop <100mV, typical.
<200mV, typical
RIPPLE AND NOISE
(20Hz to 20MHz)
3mV rms/8mV p-p, typical
REMOTE SENSE Automatic 1V max. drop in each lead. Add 2mV to the voltage load regulation specification for each 1V change in the negative output lead due to load current change

 

DC CURRENT (2 Years, 23°C ± 5°C)
OUTPUT CURRENT 0–9V: 5A max. >9V–15V: 3A max. (not intended to be operated in parallel)
SOURCE COMPLIANCE ACCURACY ±(0.16% + 5mA)
PROGRAMMED SOURCE COMPLIANCE RESOLUTION 1.25mA
READBACK ACCURACY 2303 5A range ±(0.2% + 400µA)
5mA range ±(0.2% + 1µA)
2303-PJ 5A range ±(0.2% + 400µA)
500mA range ±(0.2% + 40µA)
READBACK RESOLUTION 2303 5A range 100µA
5mA range 0.1µA
2303-PJ 5A range 100µA
500mA range 10µA
CURRENT SINK CAPACITY 0–5V: 2A max. 5V–15V: Derate 0.1A per volt above 5V
LOAD REGULATION 0.01% + 1mA
LINE REGULATION 0.5mA
STABILITY 0.01% + 50µA

 

DIGITAL VOLTMETER INPUT (2 Years, 23°C ± 5°C)
INPUT VOLTAGE RANGE 0 to +20VDC
INPUT IMPEDANCE 10Ω typical
MAXIMUM VOLTAGE (either input terminal) WITH RESPECT TO OUTPUT LOW –3V, +22V
READING ACCURACY ±(0.05% + 3mV)
READING RESOLUTION 1mV

 

DC GENERAL
MEASUREMENT TIME CHOICES 0.01 to 10 PLC , in 0.01PLC steps
AVERAGE READINGS 1 to 10
READING TIME 31ms, typical

 

PULSE CURRENT MEASUREMENT OPERATION
TRIGGER LEVEL 2303 5mA to 5A, in 5mA steps
2303-PJ 5A Range 0mA to 5A, in 5mA steps
500mA Range 0mA to 500mA, in 0.5mA steps
TRIGGER DELAY 0 to 100ms, in 10µs steps
INTERNAL TRIGGER DELAY 25µs
HIGH/LOW/AVERAGE MODE Measurement Aperture Settings 33.3µs to 833ms, in 33.3µs steps
Average Readings 1 to 100
BURST MODE Measurement Aperture 33.3µs
Conversion Rate 3600/second, typical
Number of Samples 1 to 5000
Transfer Samples Across IEEE Bus in Binary Mode 4800 bytes/second, typical
LONG INTEGRATION MODE Measurement Time 850ms (840ms) to 60 seconds in 16.7ms (20ms) steps