Đồng hồ vạn năng để bàn Keithley 2016

Đồng hồ vạn năng để bàn Keithley 2016

• Hãng sản xuất

Model

Bảo hành

Keithley

2016

12 tháng

Để có giá tốt nhất và nhanh nhất, vui lòng liên hệ

(08) 3977 8269 – (08) 3601 6797

Email: sales@lidinco.com – lidinco@gmail.com

  • Mô tả
  • Thông số kỹ thuật
  • Phụ kiện

Mô tả

2016

Đồng hồ vạn năng độ chính xác cao  Keithley 2016

Thông tin thiết bị:

  • THD, THD+Noise, and SINAD measurements
  • 20Hz�20kHz sine wave generator
  • Fast frequency sweeps
  • 2016, 2016-P: 9.5Vrms single-ended or 19Vrms differential output
  • Individual harmonic magnitude measurements
  • 5 standard audio shaping filters
  • 13 DMM functions (6-1/2 digits)

 

 

 

 

 

 


 

 

 

GENERAL SPECIFICATIONS
POWER SUPPLY 100V / 120V / 220V / 240V.
LINE FREQUENCY 50Hz to 60Hz and 400Hz, automatically sensed at power-up.
POWER CONSUMPTION 40 VA.
OPERATING ENVIRONMENT Specified for 0°C to 50°C. Specified to 80% R.H. at 35°C. RH at 35°C and at an altitude of up to 2,000 meters.
STORAGE ENVIRONMENT –40°C to 70°C.
WARRANTY 3 years.
SAFETY Conforms to European Union Directive 73/23/EEC EN61010-1.
EMC Complies with European Union Directive 89/336/EEC, EN61326-1.
VIBRATION MIL-PRF-28800F Class 3 Random.
WARMUP 1 hour to rated accuracy.
DIMENSIONS Rack Mounting: 89mm high × 213mm wide × 370mm deep (3 1 / 2 in × 8 3 / 8 in × 14 9 / 16 in).
Bench Configuration (with handle and feet) 104mm high × 238mm wide × 370mm deep (4 1 / 8 in × 9 3 / 8 in × 14 9 / 16 in).
NET WEIGHT 4.2kg (8.8 lbs).
SHIPPING WEIGHT 5kg (11 lbs).
VOLT HERTZ PRODUCT =8 × 10 7 V·Hz.
ACCESSORIES SUPPLIED Model 1751 Safety Test Leads, User Manual, Service Manual.

 

DISTORTION CHARACTERISTICS
VOLTAGE RANGE 100mV, 1V, 10V, 100V, 750V (user selectable).
INPUT IMPEDANCE 1MΩ paralleled by <100pF
DISPLAY RANGE 0–100% or 0–100.00dB
RESOLUTION 0.0001% or 0.00001dB
FUNDAMENTAL FREQUENCY RANGE 20Hz–20kHz
HARMONIC FREQUENCY RANGE 40Hz–50kHz
FREQUENCY RESOLUTION 0.008Hz
FREQUENCY ACCURACY ±0.01% of reading
FREQUENCY TEMPERATURE COEFFICIENT =100ppm over operating temperature range
MEASUREMENT MODE ACCURACY (1 Year, 23°C ±5°C) RESIDUAL DISTORTION
THD and individual harmonic magnitudes ±0.8dB, 20Hz to 20kHz 2 0.004% or –87dB
20Hz to 20kHz
THD + n ±1.5 dB,
100Hz to 20kHz 2
0.056% or –65dB
20Hz to 20kHz
SINAD ±1.5dB
100Hz to 20kHz
+65dB
20Hz to 20kHz
AC Level
V rms
±(0.13% of reading + 0.009% of range)
20Hz to 20kHz
 
External Interface
None C-Message
CCITT Weighting CCIR/ARM
CCIR “A” Weighting
NUMBER OF HARMONICS INCLUDED IN THD CALCULATION 2 to 64 (user selectable)
HI AND LO CUTOFF FILTERS (bus settable) 20Hz–50kHz. Can be combined to form brickwall bandpass filter
Distortion Measurement Reading Rate
FUNDAMENTAL FREQUENCY ACQUISITION MODE FUNDAMENTAL FREQUENCY RANGE MINIMUM READINGS PER SECOND
Single acquisition or stored value 20 Hz to 100 Hz
100 Hz to 1 kHz
1 kHz to 20 kHz
14
24
28
Automatic 20 Hz to 30 Hz
30 Hz to 400 Hz
400 Hz to 20 kHz
5.5
6
6.6
Frequency Sweep Reading Rate
NUMBER OF FREQUENCIES TIME (seconds)
5 0.2
30 1.1
100 3.5
200 6.9
Notes
  1. Input signal at full scale.
  2. V IN =20% of range and harmonics >–65dB.
  3. Speeds are for default operating conditions (*RST), and display off, auto range off, binary data transfer, trig delay = 0.
  4. Typical times: frequencies in 400–4kHz range, binary data transfer, TRIG DELAY = 0, Display OFF, Auto Range OFF. Data returned is THD measurement plus AC voltage.

 

GENERATOR CHARACTERISTICS
FREQUENCY RANGE 10–20kHz
FREQUENCY RESOLUTION 0.007Hz.
FREQUENCY ACCURACY ±(0.015% of reading + 0.007Hz)
FREQUENCY TEMPERATURE COEFFICIENT <100ppm over operating temperature range.
SOURCE OUTPUT
Waveform Sinewave
Amplitude Range 2V rms (50Ω and 600Ω) or 4V rms (HI Z)
Amplitude Resolution 0.5mV rms (50Ω and 600Ω) or 1mV rms (HI Z)
Amplitude Accuracy ±(0.3% of setting + 2mV)
Amplitude Temperature Coefficient Typically 0.015%/°C
Amplitude Flatness ±0.1dB
Output Impedance 50O ± 1Ω or 600O ± 10Ω, user selectable
THD –64dB
Noise 100µV rms
DC Offset Voltage ±1.2mV
INV/PULSE OUTPUT (SINEWAVE MODE)
Frequency Same as source output
Amplitude Range 2V rms (50Ω and 600Ω) or 4V rms (HI Z)
Amplitude Resolution 0.5mV (50Ω and 600Ω) or 1mV rms (HI Z)
Amplitude Accuracy ±(2.0% of setting + 2mV)
Amplitude Flatness ±0.1dB
Output Impedance Same as Source Output setting
THD –64dB
Noise 100µV rms
DC Offset Voltage ±1.1mV typ., ±13mV max
INV/PULSE OUTPUT (PULSE MODE)
Frequency Same as source output
Duty Cycle 45% ±3%
Output Impedance Same output impedance as the source output
Amplitude 0.0V ±0.07V to 4.9V ±0.12V pulse open circuit
0.0V ±0.05V to 3.3V ±0.08V pulse 100Ω load
Overshoot 1.0V maximum pulse open circuit
0.2V maximum with 100Ω load pulse open circuit
Undershoot 1.1V maximum pulse open circuit
0.45V maximum with 100Ω load pulse open circuit
Notes
  1. 1 year, 23°C ±5°C
  2. Measured at V OUT = 0V with gain 100 amplifier and 2-pole 50kHz low pass filter, Inv/Pulse in sinewave mode, HI Z output impedance, and no load
  3. With HI Z output impedance and 1m 50Ω coaxial cable
  4. HI Z output impedance, no load
  5. 4V output
  6. THD measurement includes harmonics 2 through 5, 1V rms output, HI Z, no load.

 

DC CHARACTERISTICS
CONDITIONS: MED (1 PLC)1 OR SLOW (10 PLC)ACCURACY: ±(PPM OF READING + PPM OF RANGE) OR MED (1 PLC) WITH FILTER OF 10(PPM = PARTS PER MILLION) (E.G., 10PPM = 0.001%)
FUNCTION RANGE RESOLUTION TEST CURRENT
OR BURDEN
VOLTAGE (±5%)
INPUT
RESISTANCE
24 HOUR
23°C ± 1°
90 DAY
23°C ± 5°
1 YEAR
23°C ± 5°
TEMPERATURE
COEFFICIENT
0°–18°C & 28°–50°C
VOLTAGE 100.0000 mV
1.000000 V
10.00000 V
100.0000 V
1000.000 V
0.1 µV
1.0 µV
10 µV
100 µV
1 mV
  > 10 GΩ
> 10 GΩ
> 10 GΩ
10 MΩ ±1%
10 MΩ ±1%
30 + 30
15 + 6
15 + 4
15 + 6
20 + 6
40+35
25+7
20+5
30+6
35+6
50+35
30+7
30+5
45+6
45+6
52 + 6
2 + 1
2 + 1
5 + 1
5 + 1
RESISTANCE 1100.0000 Ω
1.000000 kΩ
10.00000 kΩ
100.0000 kΩ
1.000000 MΩ
10.00000 MΩ
100.0000 MΩ
100 µΩ
1 mΩ
10 mΩ
100 mΩ
1 Ω
10 Ω
100 Ω
1 mA
1 mA
100 µA
10 µA
10 µA
700 nA // 10MΩ
700 nA // 10MΩ
  30 + 30
20 + 6
20 + 6
20 + 6
20 + 6
300 + 6
1600 + 30
80 + 40
80 + 10
80 + 10
80 + 10
80 + 10
450 + 10
2000 + 30
100 + 40
100 + 10
100 + 10
100 + 10
100 + 10
600 + 10
2200 + 30
8 + 6
8 + 1
8 + 1
8 + 1
8 + 1
70 + 1
385 + 1
CURRENT 10.00000 mA
100.0000 mA
1.000000 A
3.00000 A
10 nA
100 nA
1 µA
10 µA
< 0.15 V
< 0.03 V
< 0.3 V
< 1 V
  60 + 30
100+300
200+30
1000+15
300+80
300+800
500 + 80
1200 + 40
500 + 80
500 + 800
800 + 80
1200 + 40
50 + 5
50 + 50
50 + 5
50 + 5
CONTINUITY 2W 1 kΩ 100 mΩ 1 mA   40 + 100 100 + 100 120 + 100 8 + 1
DIODE TEST 3.00000 V
10.00000 V
10.00000 V
10 µV
10 µV
10 µV
1 mA
100 µA
10 µA
  20 + 6
20 + 6
20 + 6
30 + 7
30 + 7
30 + 7
40 + 7
40 + 7
40 + 7
8 + 1
8 + 1
8 + 1

 

DC OPERATING CHARACTERISTICS
FUNCTION DIGITS READINGS/s PLCs
DCV (all ranges),
DCI (all ranges), and
2W Ohms (<10M range)
6 1 / 2
6 1 / 2
6 1 / 2
5 1 / 2
5 1 / 2
5 1 / 2
4 1 / 2
5
30
50
270
500
1000
2000
10
1
1
0.1
0.1
0.04
0.01

 

DC SYSTEM SPEEDS
RANGE CHANGE 50 / s
FUNCTION CHANGE 45 / s
AUTORANGE TIME <30 ms
ASCII READINGS TO RS-232 (19.2K BAUD) 55 / s.
MAX. INTERNAL TRIGGER RATE 2000 / s.
MAX. EXTERNAL TRIGGER RATE 400 / s.
DC GENERAL
LINEARITY OF 10VDC RANGE ±(2ppm of reading + 1ppm of range)
DCV, Ω, TEMPERATURE, CONTINUITY, DIODE TEST INPUT PROTECTION 1000V, all ranges
MAXIMUM 4WΩ LEAD RESISTANCE 10% of range per lead for 100Ω and 1kΩ ranges; 1kΩ per lead for all other ranges
DC CURRENT INPUT PROTECTION 3A, 250V fuse
SHUNT RESISTOR 0.1Ω for 3A, 1A and 100mA ranges. 10Ω for 10mA range
CONTINUITY THRESHOLD Adjustable 1O to 1000Ω.
AUTOZERO OFF ERROR Add ±(2ppm of range error + 5µV) for <10 minutes and ±1°C change
OVERRANGE 120% of range except on 1000V, 3A and Diode.
DC Notes
  1. Add the following to ppm of range accuracy specification based on range:1V and 100V, 2ppm; 100mV, 15ppm; 100Ω, 15ppm; <1MΩ, 2ppm; 10mA and 1A, 10ppm; 100mA, 40ppm
  2. Speeds are for 60 Hz operation using factory default operating conditions (*RST). Autorange off, Display off, Trigger delay = 0
  3. Speeds include measurement and binary data transfer out the GPIB
  4. Auto zero off
  5. Sample count = 1024, auto zero off
  6. Auto zero off, NPLC = 0.01
  7. Ohms = 24 readings/second
  8. 1 PLC = 16.67ms @ 60Hz, 20ms @ 50Hz/400Hz. The frequency is automatically determined at power up
  9. For signal levels >500V, add 0.02ppm/V uncertainty for the portion exceeding 500V.
  10. Add 120ms for ohms
  11. Must have 10% matching of lead resistance in Input HI and LO
  12. For line frequency ±0.1%
  13. For 1kΩ unbalance in LO lead
  14. Relative to calibration accuracy
  15. Specifications are for 4-wire ohms. For 2-wire ohms, add 1Ω additional uncertainty.
  16. For rear inputs. Add the following to temperature Coefficient ‘ppm of reading’ uncertainty; 10MΩ 70ppm, 100MΩ 385ppm. Operating environment specified for 0°C to 50°C, 50% RH at 35°C

 

TRUE RMS AC VOLTAGE AND CURRENT CHARACTERISTICS
ACCURACY 1 : ±(% of reading + % of range), 23°C ±5 °C
VOLTAGE RANGE CALIBRATION
CYCLE
3 Hz–
10 Hz
10 Hz–
20 kHz
20 kHz–
50 kHz
50 kHz–
100 kHz
100 kHz–
300 kHz
100.0000 mV
1.000000 V
10.00000 V
100.0000 V
750.000 V
0.1 µV
1.0 µV
10 µV
100 µV
1 mV
90 Days1 Year 0.35+0.030.35+0.03 0.05+0.030.06+0.03 0.11+0.050.12+0.05 0.60+0.080.60+0.08 4 + 0.54 + 0.5
  TEMPERATURE
COEFFICIENT/°C
0.035+0.003 0.005+0.003 0.006+0.005 0.01+0.006 0.03+0.01
CURRENT
RANGE
RESOLUTION CALIBRATION
CYCLE
3 Hz–
10 Hz
10 Hz–
3 kHz
3 kHz-
5 kHz
 
1.000000 A 1 µA 90 Day/1 Year 0.30+0.04 0.10+0.04 0.14+0.04  
3.00000 A 10 µA 90 Day/1 Year 0.35+0.06 0.15+0.06 0.18+0.06  
  TEMPERATURE
COEFFICIENT/°C
0.035+0.006 0.015+0.006 0.015+0.006  

 

HIGH CREST FACTOR ADDITIONAL ERROR ±(% of reading)
CREST FACTOR 1–2 2–3 3–4 4–5
ADDITIONAL ERROR 0.05 0.15 0.30 0.40
AC OPERATING CHARACTERISTICS
FUNCTION DIGITS READINGS/s RATE BANDWIDTH
ACV (all ranges) 6 1 / 2 2s/reading SLOW 3 Hz–300 kHz
ACI (all ranges) 6 1 / 2 1.4 MED 30 Hz–300 kHz
  6 1 / 2 4.8 MED 30 Hz–300 kHz
  6 1 / 2 2.2 FAST 300 Hz–300 kHz
  6 1 / 2 35 FAST 300 Hz–300 kHz
ADDITIONAL LOW FREQUENCY ERRORS ±(% of reading)
FUNCTION SLOW MED FAST
20Hz – 30Hz 0 0.3 -
30Hz – 50Hz 0 0 -
50Hz – 100Hz 0 0 1.0
100Hz – 200Hz 0 0 0.18
200Hz – 300Hz 0 0 0.10
> 300Hz 0 0 0
AC SYSTEM SPEEDS
FUNCTION/RANGE CHANGE 4 / s
AUTORANGE TIME <3 s
ASCII READINGS TO RS-232 (19.2K BAUD) 50 / s
MAX. INTERNAL TRIGGER RATE 300 / s
MAX. EXTERNAL TRIGGER RATE 260 / s
AC GENERAL
INPUT IMPEDANCE 1MΩ ±2% paralleled by <100pF
ACV INPUT PROTECTION 1000Vp
MAXIMUM DCV 400V on any ACV range
ACI INPUT PROTECTION 3A, 250V fuse
BURDEN VOLTAGE 1A Range: <0.3V rms. 3A Range: <1V rms
SHUNT RESISTOR 0.1Ω on all ACI ranges
AC CMRR >70dB with 1kΩ in LΩ lead
MAXIMUM CREST FACTOR 5 at full scale
VOLT HERTZ PRODUCT =8 × 10 7 V·Hz
OVERRANGE 120% of range except on 750V and 3A ranges
AC Notes
  1. Specifications are for SLOW rate and sinewave inputs >5% of range
  2. Speeds are for 60 Hz operation using factory default operating conditions (*RST). Auto zero off, Auto range off, Display off, includes measurement and binary data transfer out the GPIB
  3. 0.01% of step settling error. Trigger delay = 400ms
  4. Trigger delay = 0
  5. DETector:BANDwidth 300, NPLC = 0.01
  6. Maximum useful limit with trigger delay = 175ms
  7. Applies to non-sine waves >5Hz and <500Hz. (Guaranteed by design for Crest Factors >4.3)
  8. Applies to 0°–18°C and 28°–50°C
  9. For signal levels >2.2A, add additional 0.4% to “of reading” uncertainty
  10. Typical uncertainties. Typical represents two sigma or 95% of manufactured units measure < 0.35% of reading and three sigma or 99.7% < 1.06% of reading.
TRIGGERING AND MEMORY
READING HOLD SENSITIVITY 0.01%, 0.1%, 1%, or 10% of reading
TRIGGER DELAY 0 to 99 hrs (1ms step size)
EXTERNAL TRIGGER LATENCY 200µs + <300µs jitter with autozero off, trigger delay = 0
MEMORY 1024 readings
MATH FUNCTIONS
Rel, Min/Max/Average/StdDev (of stored reading), dB, dBm, Limit Test, %, and mX+b with user defined units displayed.
dBm REFERENCE RESISTANCES 1 to 9999O in 1Ω increments
STANDARD PROGRAMMING LANGUAGES
SCPI (Standard Commands for Programmable Instruments)
REMOTE INTERFACE
GPIB (IEEE-488.1, IEEE-488.2) and RS-232C
FREQUENCY AND PERIOD CHARACTERISTICS
ACV
RANGE
FREQUENCY
RANGE
PERIOD
RANGE
GATE
TIME
RESOLUTION
±(ppm of
reading)
ACCURACY
90 DAY/1 YEAR
±(% of reading)
100 mV
to
750 V
3 Hz
to
500 kHz
333 ms
to
2 µs
1 s (SLOW)
0.1 s (MED)
10 ms (FAST)
0.333
3.33
33.3
0.01
0.01
0.01
Frequency Notes
  1. Specifications are for square wave inputs only. Input signal must be >10% of ACV range. If input is <20mV on the 100mV range then the frequency must be >10Hz
  2. 20% overrange on all ranges except 750V range
TEMPERATURE CHARACTERISTICS
TYPE RANGE RESOLUTION 90 DAY/1 YEAR (23°C ± 5°C)
ACCURACY
Relative to
Reference Junction
J –200 to+760°C 0.001°C ±0.5°C
K –200 to + 1372°C 0.001°C ±0.5°C
T –200 to + 400°C 0.001°C ±0.5°C

Keithley 2016 Product Specifications Summary:
(for full specs please see the full specs tab or the data sheet)

Multimeters Template

Type Digital
Style Bench
Measures AC V Yes
Measures DC V Yes
Measures AC A Yes
Measures DC A Yes
True RMS What's This? Yes
Backlight No
Manual/Autoranging What's This? Auto-ranging
Analog Bargraph on Display No

Test Equipment General Attributes

Calibration Included Factory Calibration
Power Supply Voltage 120V 50/60 Hz
  • 4288-1 Rack Mount Kit
    Single-unit rack mount kit. Height: 88mm (3-1/2 in)
  • 4288-2 Rack Mount Kit
    Dual-unit rack mount kit. Height: 88mm (3-1/2 in) 
  • 7009-5 RS-232 Cable
    Shielded
  • 8501-1 Trigger Link Cable
    With Male Micro-DIN Connectors at each End, 1m (3.3 ft)
  • 8501-2 Trigger Link Cable
    With Male Micro-DIN Connectors at each End, 2m (6.6 ft)
  • 8502 Micro-DIN to 6 BNCs Adapter Box
    Discontinued. 
  • 8503 DIN-to-BNC Trigger Cable
    2 Male BNCs, 1m (3.2 ft.)