System is used for vibration calibration Spektra CS Q-LEAP™ SINE with SE-29
Hotline: +84 906 988 447
Head Office: Ho Chi Minh City
- Tel: +84 2839 778 269 / 3601 6797
- Email: sales@lidinco.com
- Add: 487 Cong Hoa Street, Ward 15, Tan Binh Dist, HCM City, Vietnam
Office: Bac Ninh City
- Tel: +84 222 730 0180
- Email: bn@lidinco.com
- Add: 184 Binh Than Street, Vo Cuong Ward, Bac Ninh City, Vietnam
- Technical Counseling 100% Free
- Free Shipping For 3.000.000vnd Order
Specification
Data is being updated
The CS Q-LEAP™ system is used for vibration calibration of a wide range of sensors including vibration sensors and measurement systems. It is based on the HERO™ vibration controller and the SE-29 high-frequency vibration exciter.
Key features
- Frequency range: 3 Hz...20 kHz (50 kHz)
- Traceable to PTB (German National Metrology Laboratory)
- Calibration of vibration sensors, measurement systems and calibrators
- Integrated sensor database
- Integrated e-Cal™ software for the generation of calibration certificates (print, PDF...)
- Easy data exchange with applications like ERP systems or measuring equipment databases
Technical Data
- Acceleration, max.: 450 m/s² (46 gn) peak
- Force rating, max.: 95 N (21 lbf) peak
- Displacement, max.: 10 mm (0.39 inch)
- Payload, max.: 2 kg (4.4 lbs) vertical / 1 kg (2.2 lbs) horizontal
CS Q-LEAP™ Software
- sine calibration
- sine sweep
- vibration measurement
- vibration generation
- more on demand
Typical DUTs
- vibration sensors
· PE transducer
· IEPE transducer
· VC transducer
· PR transducer
· Digital transducer (SPI, I2C, DTI, and other interfaces) - vibration meters
- vibration calibrators
- supports TEDS / ID modules according to IEEE 1451.4
Standards
- ISO 16063 - 21: Calibration of vibration transducers by comparison to a reference transducer
- ISO 17025: General requirements for the competence of testing and calibration laboratories
- ISO 16063 - 41: Calibration of laser vibrometers