Contacting Extensometers Ametek STGA Series
Hotline: +84 906 988 447
Head Office: Ho Chi Minh City
- Tel: +84 2839 778 269 / 3601 6797
- Email: sales@lidinco.com
- Add: 487 Cong Hoa Street, Ward 15, Tan Binh Dist, HCM City, Vietnam
Office: Bac Ninh City
- Tel: +84 222 730 0180
- Email: bn@lidinco.com
- Add: 184 Binh Than Street, Vo Cuong Ward, Bac Ninh City, Vietnam
- Technical Counseling 100% Free
- Free Shipping For 3.000.000vnd Order
Measurement Principle: Full bridge strain gauge
Accuracy: 0.5%
Maximum Clamp Opening: 25 mm (0.98 in)
Maximum Elongation: 10%, 20%, 25%
Selectable Gauge Lengths: 10mm (0.4 in)
25 mm (1 in)
50 mm (2 in)
80 mm (3.2 in)
Temperature Range: -40° C to +100° C
Excitation: 10 VDC typical
Output: 2-4 mV/V nominal
Description Ametek STGA series
We provide a wide variety of contacting extensometers in just the size you need.
Our general purpose, precision extensometers are designed for testing a wide range of materials including metals, plastics, composites and ceramics operating in tension, compression and cyclic testing modes.
Its dual flexure design means a higher frequency operation is permitted, while eliminating vibration sensitivity at the same time. Each extensometer is supplied with an attachment kit to allow quick one-hand mounting to the sample.
Contacting Extensometers Key Features:
- Available in many sizes
- Easy mounting to the sample
For more information about our range of contacting extensometers please see the Epsilon Extensometer Catalogue.
Request more information to see how we can help you with a materials testing solution that suits your need.
Features STGA series
• Tests specimen through failure
• Suitable for axial tension, compression and cyclic testing
• Overtravel safety stop eliminates risk of damage
• Interchangeable components: arms, spacers and steel knife edges
• Meets ASTM E83 Class B-1, ISO 9513, Class 0.5
• Simple one hand mounting to sample
Specifications
Model | Part No | |
High Resolution Strain Gauge Extensometers for use with LR Series and LS100 Series Machines: | ||
STGA10/10/L | 10 mm (0.39 in) gauge length 10% max elongation | 01/2793 |
STGA25/10/L | 25 mm (0.98 in) gauge length 10% max elongation | 01/2794 |
STGA25/25/L | 25 mm (0.98 in) gauge length 25% max elongation | 01/2795 |
STGA25/50/L | 25 mm (0.98 in) gauge length 50% max elongation | 01/2796 |
STGA25/100/L | 25 mm (0.98 in) gauge length 100% max elongation | 01/2797 |
STGA50/25/L | 50 mm (1.97 in) gauge length 25% max elongation | 01/2798 |
STGA50/50/L | 50 mm (1.97 in) gauge length 50% max elongation | 01/2799 |
STGA50/100/L | 50 mm (1.97 in) gauge length 100% max elongation | 01/2800 |
STGA80/10/L | 80 mm (3.15 in) gauge length 10% max elongation | 01/2801 |
STGA80/25/L | 80 mm (3.15 in) gauge length 25% max elongation | 01/2802 |
Model | Part No | |
High Resolution Strain Gauge Extensometers for use with EZ and Plus Series Machines: | ||
STGA10/10/E | 10 mm (0.39 in) gauge length 10% max elongation | 01/2917 |
STGA25/10/E | 25 mm (0.98 in) gauge length 10% max elongation | 01/2918 |
STGA25/25/E | 25 mm (0.98 in) gauge length 25% max elongation | 01/2919 |
STGA25/50/E | 25 mm (0.98 in) gauge length 50% max elongation | 01/2920 |
STGA25/100/E | 25 mm (0.98 in) gauge length 100% max elongation | 01/2921 |
STGA50/25/E | 50 mm (1.97 in) gauge length 25% max elongation | 01/2922 |
STGA50/50/E | 50 mm (1.97 in) gauge length 50% max elongation | 01/2923 |
STGA50/100/E | 50 mm (1.97 in) gauge length 00% max elongation | 01/2924 |
STGA80/10/E | 80 mm (3.15 in) gauge length 10% max elongation | 01/2925 |
STGA80/25/E | 80 mm (3.15 in) gauge length 25% max elongation | 01/2926 |
Accessories
Please login to write review!