Description
The 7½-digit 2010 Series Low Noise Multimeter combines cost effectiveness and high resolution with the throughput and accuracy needed for high speed DMM applications such as testing precision sensors, transducers, A/D and D/A converters, regulators, references, connectors, switches, and relays. Based on the same high speed, low noise A/D converter technology as the 2000, 2001, and 2002 Series, the 2010 Series adds four unique resistance measurement functions that make it ideal for characterizing the resistance, linearity, or isolation of contacts, connectors, switches, or relays.
Features:
- 7-1/2 digit resolution
- 100nV rms noise floor
- 7ppm DCV repeatability
- Built-in 10-channel scanner mainframe
- Dry circuit and low power measurement mode
- 15 measurement functions including support for RTD and thermocouple temperature measurements
- Built-in ratio measurement function
- GPIB and RS-232 interfaces
General specifications | ||
POWER SUPPLY | 100V / 120V / 220V / 240V | |
LINE FREQUENCY | 50Hz to 60Hz and 440Hz, automatically sensed at power-up | |
POWER CONSUMPTION | 22VA | |
VOLT HERTZ PRODUCT | =8 × 10 7 V·Hz | |
OPERATING ENVIRONMENT | Specified for 0° to 50°C. Specified to 80% R.H. at 35°C | |
STORAGE ENVIRONMENT | –40° to 70°C | |
ALTITUDE | Up to 2000 meters | |
SAFETY | Conforms to European Union Directive 73/23/EEC EN 61010-1, Cat II | |
EMC | Complies with European Union Directive 89/336/EEC, EN 61326-1 | |
VIBRATION | MIL-PRF-28800F Class 3 Random | |
WARMUP | 2 hours to rated accuracy | |
DIMENSIONS | Rack Mounting | 89mm high × 213mm wide × 370mm deep (3½ in × 8 3 / 8 in × 14 9 / 16 in) |
Bench Configuration (with handle and feet) | 104mm high × 238mm wide × 370mm deep (4 1 / 8 in × 9 3 / 8 in × 14 9 / 16 in) | |
SHIPPING WEIGHT | 5kg (11 lbs) |
DC VOLTAGE | ||||
Range | Resolution | Accuracy 23°C ± 5°C ±(ppm of rdg. + ppm of range) | Input Resistance | |
90 Day | 1 Year | |||
100.00000 mV | 10 nV | 25 + 9 | 37 + 9 | > 10 GΩ |
1.0000000 V | 100 nV | 18 + 2 | 25 + 2 | > 10 GΩ |
10.000000 V | 1 µV | 18 + 4 | 24 + 4 | > 10 GΩ |
100.00000 V | 10 µV | 25 + 5 | 35 + 5 | 10 MΩ ±1% |
1000.0000 V | 100 µV | 31 + 6 | 41 + 6 | 10 MΩ ±1% |
RESISTANCE | ||||
Range | Resolution | Accuracy 23°C ± 5°C ±(ppm of rdg. + ppm of range) | Test Current | |
90 Day | 1 Year | |||
10.000000 Ω | 1 µΩ | 40 + 9 | 60 + 9 | 10 mA |
100.00000 Ω | 10 µΩ | 36 + 9 | 52 + 9 | 1 mA |
1.0000000 kΩ | 100 µΩ | 33 + 2 | 50 + 2 | 1 mA |
10.000000 kΩ | 1 mΩ | 32 + 2 | 50 + 2 | 100 µA |
100.00000 kΩ | 10 mΩ | 40 + 4 | 70 + 4 | 10 µA |
1.0000000 MΩ | 100 mΩ | 50 + 4 | 70 + 4 | 10 µA |
10.000000 MΩ | 1 Ω | 200 + 4 | 400 + 4 | 640 nA//10 MΩ |
100.00000 MΩ | 10 Ω | 1500 + 4 | 1500 + 4 | 640 nA//10 MΩ |
DC CURRENT | ||||
Range | Resolution | Accuracy 23°C ± 5°C ±(ppm of rdg. + ppm of range) | Burden Voltage | |
90 Day | 1 Year | |||
10.000000 mA | 1 nA | 300 + 80 | 500 + 80 | < 0.15 V |
100.00000 mA | 10 nA | 300 + 800 | 500 + 800 | < 0.18 V |
1.0000000 A | 100 nA | 500 + 80 | 800 + 80 | < 0.35 V |
3.000000 A | 1 µA | 1200 + 40 | 1200 + 40 | < 1 V |
CONTINUITY 2W | ||||
Range | Resolution | Accuracy 23°C ± 5°C ±(ppm of rdg. + ppm of range) | Test Current | |
90 Day | 1 Year | |||
1 kΩ | 100 mΩ | 100 + 100 | 120 + 100 | 1 mA |
DIODE TEST | ||||
Range | Resolution | Accuracy 23°C ± 5°C ±(ppm of rdg. + ppm of range) | Test Current | |
90 Day | 1 Year | |||
10.000000 V | 1 µV | 30 + 7 | 40 + 7 | 1 mA |
4.400000 V | 1 µV | 30 + 7 | 40 + 7 | 100 µA |
10.000000 V | 1 µV | 30 + 7 | 40 + 7 | 10 µA |
DC OPERATING CHARACTERISTICS | |||
Function | Digits | Readings/s | PLCs |
10.000000 V | 7½ | 4 (3) | 5 |
6½ | 30 (27) | 1 | |
6½ | 50 (44) | 1 | |
5½ | 260 (220) | 0.1 | |
5½ | 490 (440) | 0.1 | |
5½ | 1000 (1000) | 0.04 | |
4½ | 2000 (1800) | 0.01 |
DC NOISE PERFORMANCE | |||||
Rate | Digits | RMS Noise 100mV Range (2 min.) | RMS Noise 10V Range (2 min.) | NMRR | CMRR |
5 PLC | 7½ | 110 nV | 1.2 µV | 60 dB | 140 dB |
1 PLC | 6½ | 125 nV | 1.4 µV | 60 dB | 140 dB |
0.1 PLC | 5½ | 1.9 µV | 11.5 µV | 80 dB | |
0.01 PLC | 4½ | 2.9 µV | 139 µV | 80 dB |
TRUE RMS AC VOLTAGE AND CURRENT CHARACTERISTICS | |||
Voltage Range | Resolution | Frequency Range | Accuracy (1 Year) ±(% of reading + % of range) |
100 mV to 750 V | 0.1 µV to 1 mV | 3 Hz–10 Hz | 0.35 + 0.03 |
10 Hz–20 kHz | 0.06 + 0.03 | ||
20 kHz–50 kHz | 0.12 + 0.05 | ||
50 kHz–100 kHz | 0.60 + 0.08 | ||
100 kHz–300 kHz | 4 + 0.5 |
AC OPERATING CHARACTERISTICS | ||||
Function | Digits | Readings/s | Rate | Bandwidth |
ACV (all ranges), and ACI (all ranges) | 6½ | 0.5 (0.4) | SLOW | 3 Hz–300 kHz |
6½ | 1.4 (1.5) | MED | 30 Hz–300 kHz | |
6½ | 4.0 (4.3) | MED | 30 Hz–300 kHz | |
6½ | 2.2 (2.3) | FAST | 300 Hz–300 kHz | |
6½ | 35 (30) | FAST | 300 Hz–300 kHz |
FREQUENCY AND PERIOD CHARACTERISTICS | |||||
ACV Range | Frequency Range | Period Range | Gate Time | Resolution ±(ppm of reading) | Accuracy 90 Day/1 Year ±(% of reading) |
100 mV to 750 V | 3 Hz to 500 kHz | 333 ms to 2 µs | 1 s | 0.3 | 0.01 |
TEMPERATURE CHARACTERISTICS | ||||
Thermocouple | ||||
Type | Range | Resolution | Accuracy 90 Day/1 Year (23°C ± 5°C) Relative to Reference Junction | USING 2001-TCSCAN |
J | –200 to + 760°C | 0.001°C | ±0.5°C | ±0.65°C |
K | –200 to + 1372°C | 0.001°C | ±0.5°C | ±0.70°C |
N | –200 to + 1300°C | 0.001°C | ±0.5°C | ±0.70°C |
T | –200 to + 400°C | 0.001°C | ±0.5°C | ±0.68°C |
4-WIRE RTD | ||||
Range | Resolution | Accuracy 90 Day/1 Year (23°C ± 5°C) | Accuracy 2 Years (23°C ± 5°C) | |
–100° to +100°C | 0.001°C | ±0.08°C | ±0.12°C | |
–200° to +630°C | 0.001°C | ±0.14°C | ±0.18°C |
Keithley 2010 Product Specifications Summary: (for full specs please see the full specs tab or the data sheet)
Multimeters Template | |
---|---|
Type | Digital |
Style | Bench |
Measures AC V | Yes |
Measures DC V | Yes |
Measures AC A | Yes |
Measures DC A | Yes |
True RMS What's This? | Yes |
Backlight | No |
Manual/Autoranging What's This? | Auto-ranging |
Analog Bargraph on Display | No |
Test Equipment General Attributes | |
Calibration Included | Factory Calibration |
Power Supply Voltage | 120V 50/60 Hz |
Accessories
- Model 1751 Safety Test Leads
- User Manual
- Service Manual