The 2001 Series offers 7 ½ -digit resolution and basic DCV accuracy of 0.0018%, offering exceptional resolution, accuracy, and sensitivity. This high performance digital multimeter not only delivers performance specifications usually associated only with instruments that cost thousands more, but it also offers a broad range of functions not typically available in DMMs. For greater flexibility, a built-in card slot on the rear panel allows an additional multiplexer card for multi-point measurement applications.
Features:
- True 7-1/2 (Model 2001) or 8-1/2 digit (Model 2002) resolution - Exceptional measurement integrity with high speed - High speed function and range changing - Broad range of built-in measurement functions - Multiple measurement display - Built-in 10 channel scanner option - GPIB interface - HP3458A emulation mode (Model 2002)
2001 Condensed Specifications
DC VOLTSDCV INPUT CHARACTERISTICS AND ACCURACY
Range
FullScale
Resolution
DefaultResolution
InputResistance
Accuracy ±(ppm of reading + ppm of range)
5Minutes
24Hours
90Days
1Year
2Years
200mV
±210.00000mV
10nV
100nV
>10GΩ
3+3
10+6
25+6
37+6
50+6
2V
±2.1000000V
100nV
1µV
>10GΩ
2+1.5
7+2
18+2
25+2
32+2
20V
±21.000000V
1µV
10µV
>10GΩ
2+1.5
7+4
18+4
24+4
32+4
200V
±210.00000V
10µV
100µV
10MΩ±1%
2+1.5
13+3
27+3
38+3
52+3
1000V
±1100.0000V
100µV
1mV
10MΩ±1%
10+1.5
17+6
31+6
41+6
55+6
DC VOLTS NOTES
For TCAL ±1°C, following 55-minute warm-up. TCAL is ambient temperature at calibration, which is 23°C from factory
For TCAL ±5°C, following 55-minute warm-up. Specifications include factory traceability to US NIST
When properly zeroed using REL function
DCV Transfer Stability typical applications are standard cell comparisons and relative accuracy measurements. Specs apply for 10 power line cycles, 20-reading digital filter, autozero on with type synchronous, fixed range following 2-hour warm-up at full scale to 10% of full scale, at TREF ±1°C (TREF is the initial ambient temperature). Specifications on the 1000V range are for measurements within 5% of the initial measurement value and following measurement settling
2001 AC VOLTS
Range
Normal Mode RMS90 Days, ±2°C from last AC self-cal for 1% to 100% of range±(% of reading + % of range)
20–50Hz
50–100Hz
0.1–2kHz
2–10kHz
10–30kHz
200 mV
0.25 + 0.015
0.07 + 0.015
0.03 + 0.015
0.03 + 0.015
0.035 + 0.015
2 V
0.25 + 0.015
0.07 + 0.015
0.03 + 0.015
0.03 + 0.015
0.035 + 0.015
20 V
0.25 + 0.015
0.07 + 0.015
0.04 + 0.015
0.06 + 0.015
0.08 + 0.015
200 V
0.25 + 0.015
0.07 + 0.015
0.04 + 0.015
0.06 + 0.015
0.08 + 0.015
750 V
0.25 + 0.015
0.1 + 0.015
0.08 + 0.015
0.09 + 0.015
0.12 + 0.015
AC VOLTS NOTES
Specifications apply for sinewave input, AC + DC coupling, 1 power line cycle, digital filter off, following 55 minute warm-up
For 1% to 5% of range below 750V range, and for 1% to 7% of 750V range, add 0.01% to range uncertainty. For inputs from 200kHz to 2MHz, specifications apply above 10% of range
Normal Mode RMS90 Days, ±2°C from last AC self-cal for 1% to 100% of range±(% of reading + % of range)
30–50kHz
50–100kHz
100–200kHz
0.2–1MHz
1–2MHz
200 mV
0.05 + 0.015
0.3 + 0.015
0.75 + 0.025
2 + 0.1
5 + 0.2
2 V
0.05 + 0.015
0.3 + 0.015
0.75 + 0.025
2 + 0.1
5 + 0.2
20 V
0.1 + 0.015
0.3 + 0.015
0.75 + 0.025
4 + 0.2
7 + 0.2
200 V
0.1 + 0.015
0.3 + 0.015
0.75 + 0.025
4 + 0.2
750 V
0.15 + 0.015
0.5 + 0.015
AC VOLTS NOTES
Specifications apply for sinewave input, AC + DC coupling, 1 power line cycle, digital filter off, following 55 minute warm-up
For 1% to 5% of range below 750V range, and for 1% to 7% of 750V range, add 0.01% to range uncertainty. For inputs from 200kHz to 2MHz, specifications apply above 10% of range
2001 OHMSTWO-WIRE AND FOUR-WIRE OHMS (2W and 4W Ohms Functions)
Range
FullScale
Resolution
DefaultResolution
CurrentSource
Resistance Accuracy ±(ppm of reading + ppm of range)
24Hours
90Days
1Year
2Years
20 Ω
21.000000Ω
1 µΩ
10 µΩ
9.2 mA
29+7
52+7
72+7
110+7
200 Ω
210.00000Ω
10 µΩ
100 µΩ
0.98 mA
24+7
36+7
56+7
90+7
2 kΩ
2100.0000kΩ
100 µΩ
1 mΩ
0.98 mA
22+4
33+4
50+4
80+4.5
20 kΩ
21.000000kΩ
1 mΩ
10 mΩ
89µA
19+4
32+4
50+4
80+4.5
200 kΩ
210.00000kΩ
10 mΩ
100 mΩ
7µA
20+4.5
72+4.5
90+4.5
130+5
2 MΩ
2.1000000MΩ
100 mΩ
1 Ω
770 nA
50+4.5
110+4.5
160+4.5
230+5
20 MΩ
21.000000MΩ
1 Ω
10 Ω
70 nA
160+4.5
560+4.5
900+4.5
1100+5
200 MΩ
210.00000MΩ
10 Ω
100 Ω
4.4 nA
3000 +100
10000 +100
20000 +100
30000 +100
1 GΩ
1.0500000GΩ
100 Ω
1 kΩ
4.4 nA
9000 +100
20000 +100
40000 +100
60000 +100
OHMS NOTES
Current source is typically ±9% absolute accuracy
For 2-wire mode
Specifications are for 1 power line cycle, 10 reading digital filter, Auto Zero on, 4-wire mode, offset compensation on (for 20Ω to 20kΩ ranges)
For TCAL ±1°C, following 55 minute warm-up. TCAL is ambient temperature at calibration (23°C at the factory)
For TCAL ±5°C, following 55-minute warm-up. Specifications include traceability to US NIST
When measuring resistance of inductive loads, the inductance of that load must be 10mH or less
2001 DC AMPSDCI INPUT CHARACTERISTICS AND ACCURACY
Range
FullScale
Resolution
DefaultResolution
MaximumBurdenVoltage
Accuracy ±(ppm of reading + ppm of range)
24Hours
90Days
1Year
2Years
200 µA
210.00000µA
10 pA
100 pA
0.25 V
63+25
300+25
500+25
1350+25
2 mA
2.1000000mA
100 pA
1 nA
0.31 V
64+20
300+20
400+20
750+20
20 mA
21.000000mA
1 nA
10 nA
0.4 V
65+20
300+20
400+20
750+20
200 mA
210.00000mA
10 nA
100 nA
0.5 V
96+20
300+20
500+20
750+20
2 A
2.1000000A
100 nA
1 µA
1.5 V
500+20
600+20
900+20
1350+20
DC AMPS NOTES
Specifications are for 1 power line cycle, Auto Zero on, 10 reading digital filter
For TCAL ±1°C, following 55 minute warm-up
For TCAL ±5°C, following 55 minute warm-up. Specifications include traceability to US NIST
Add 50 ppm of range for current above 0.5A for self heating
Actual maximum voltage burden = (maximum voltage burden) × (IMEASURED/IFULL SCALE)
2001 AC AMPS
Range
ACI ACCURACY90 Days, 1 Year or 2 Years, TCAL ±5°C, for 5% to 100% of range, ±(% of reading + % of range)
20Hz–50Hz
50Hz–200Hz
200Hz–1kHz
1kHz–10kHz
10kHz–30kHz
30kHz–50kHz
50kHz–100kHz
200 µA
0.35+0.015
0.2+0.015
0.4+0.015
0.5+0.015
2 mA
0.3+0.015
0.15+0.015
0.12+0.015
0.12+0.015
0.25+0.015
0.3+0.015
0.5+0.015
20 mA
0.3+0.015
0.15+0.015
0.12+0.015
0.12+0.015
0.25+0.015
0.3+0.015
0.5+0.015
200 mA
0.3+0.015
0.15+0.015
0.12+0.015
0.15+0.015
0.5+0.015
1+0.015
3+0.015
2 A
0.35+0.015
0.2+0.015
0.3+0.015
0.45+0.015
1.5+0.015
4+0.015
AC AMPS NOTES
Specifications apply for sinewave input, AC+DC coupling, 1 power line cycle, digital filter off, following 55 minute warm-up
Add 0.005% of range uncertainty for current above 0.5A rms for self-heating
Typical values
FREQUENCY COUNTER
AC VOLTAGE INPUT
1Hz–15MHz
ACCURACY
±(0.03% of reading)
DC IN-CIRCUIT CURRENT
TYPICAL RANGES
Current
100µA to 12A
Trace Resistance
1mΩ to 10Ω typical
TEMPERATURE
Built-in linearization for J, K, N, T, E, R, S, B thermocouple types to ITS-90 and 100Ω platinum RTDs DIN 43 760 or IPTS-68
2002 General Specification
POWER
Voltage
90–134V and 180–250V, universal self-selecting
Frequency
50Hz, 60Hz, or 400Hz self-identifying at power-up.
Consumption
<55VA
ENVIRONMENTAL
Operating Temperature
0° to 50°C
Storage Temperature
–40° to 70°C
Humidity
80% R.H., 0° to 35°C
PHYSICAL
Case Dimensions
90mm high × 214mm wide × 369mm deep (3½ in. × 8½ in. × 14½ in.)
Net Weight
<4.2kg (<9.2 lbs.)
Shipping Weight
<9.1kg (<20 lbs.)
STANDARDS
EMI/RFI
Conforms to European Union EMC directive
Safety
Conforms to European Union Low Voltage directive
Note 1
For MIL-T-28800E, applies to Type III, Class 5, Style E
2002 Condensed Specifications
DC VOLTSDCV INPUT CHARACTERISTICS AND ACCURACY
Enhanced Accuracy – 10PLC, DFILT 10
Range
FullScale
Resolution
InputResistance
Accuracy ±(ppm of reading + ppm of range)
Transfer
24Hours
90Days
1Year
2Years
200mV
±210.00000mV
1 nV
>100 GΩ
0.4+1.5
3.5+3
15+8
19+9
23+10
2V
±2.1000000V
10nV
>100GΩ
0.2+0.15
1.2+0.3
6+0.8
10+0.9
14+1
20V
±21.000000V
100nV
>100GΩ
0.1+0.05
1.2+0.1
6+0.15
10+0.15
14+0.15
200V
±210.00000V
1µV
10MΩ±1%
0.5+0.08
5+0.4
14+2
22+2
30+2
1000V
±1100.0000V
10µV
10MΩ±1%
1+0.05
5+0.08
14+0.4
22+0.4
30+0.4
Normal Accuracy – 1PLC, DFILT off
Range
FullScale
Resolution
InputResistance
Accuracy ±(ppm of reading + ppm of range)
24Hours
90Days
1Year
2Years
200mV
±210.00000mV
10 nV
>100 GΩ
3.5+6
15+11
19+12
23+13
2V
±2.1000000V
100nV
>100GΩ
1.2+0.6
6+1.1
10+1.2
14+1.3
20V
±21.000000V
100nV
>100GΩ
3.2+0.35
8+0.4
12+0.4
16+0.4
200V
±210.00000V
10µV
10MΩ±1%
5+1.2
14+2.8
22+2.8
30+2.8
1000V
±1100.0000V
100µV
10MΩ±1%
5+0.04
14+0.7
22+0.7
30+0.7
DC VOLTS NOTES
Specifications are for 10 power line cycles, synchronous autozero, 10-reading repeat digital filter, autorange off, except as noted
For TCAL ±1°C, following 4-hour warm-up. TCAL is ambient temperature at calibration (23°C at the factory). Add 0.5ppm of reading uncertainty if the unit is power cycled during this interval
For TCAL ±5°C, following 4-hour warm-up
Care must be taken to minimize thermal offsets due to operator cables
Specifications apply for 20-reading repeat digital filter, TREF ±0.5°C (TREF is the initial ambient temperature), and for measurements within 10% of the initial measurement value and within 10 minutes of the initial measurement time.
Add 20ppm × (VIN/1000V) 2 additional uncertainty for inputs above 200V, except in transfer accuracy specifications
Specifications are for 1 power line cycle, normal autozero, digital filter off, autorange off.
2002 AC VOLTS Normal Mode RMS
Range
Normal Mode RMS90 Days, ±2°C from last AC self-cal for 1% to 100% of range±(% of reading + % of range)
20–50Hz
50–100Hz
0.1–2kHz
2–10kHz
10–30kHz
200 mV
0.25 + 0.015
0.07 + 0.015
0.02 + 0.02
0.02 + 0.02
0.025 + 0.02
2 V
0.25 + 0.015
0.07 + 0.015
0.02 + 0.02
0.02 + 0.02
0.025 + 0.02
20 V
0.25 + 0.015
0.07 + 0.015
0.03 + 0.015
0.04 + 0.015
0.05 + 0.015
200 V
0.25 + 0.015
0.07 + 0.015
0.03 + 0.015
0.04 + 0.015
0.05 + 0.015
750 V
0.25 + 0.015
0.1 + 0.015
0.05 + 0.015
0.06 + 0.015
0.08 + 0.015
AC VOLTS NOTES
Specifications apply for sinewave input, AC + DC coupling, 1 power line cycle, autozero on, digital filter off, following 55-minute warm-up
For 1% to 5% of range below 750V range, and for 1% to 7% of 750V range, add 0.01% of range uncertainty. For inputs from 200kHz to 2MHz, specifications apply above 10% of range.
Add 0.001% of reading × (VIN/100V) 2 additional uncertainty for inputs above 100V rms.
Typical values
2002 AC VOLTS
Range
Normal Mode RMS90 Days, ±2°C from last AC self-cal for 1% to 100% of range±(% of reading + % of range)
30–50kHz
50–100kHz
100–200kHz
0.2–1MHz
1–2MHz
200 mV
0.05 + 0.01
0.3 + 0.015
0.75 + 0.025
2 + 0.1
5 + 0.2
2 V
0.05 + 0.01
0.3 + 0.015
0.75 + 0.025
2 + 0.1
5 + 0.2
20 V
0.07 + 0.015
0.3 + 0.015
0.75 + 0.025
4 + 0.2
7 + 0.2
200 V
0.07 + 0.015
0.3 + 0.015
0.75 + 0.025
4 + 0.2
750 V
0.1 + 0.015
0.5 + 0.015
AC VOLTS NOTES
Specifications apply for sinewave input, AC + DC coupling, 1 power line cycle, autozero on, digital filter off, following 55-minute warm-up
For 1% to 5% of range below 750V range, and for 1% to 7% of 750V range, add 0.01% of range uncertainty. For inputs from 200kHz to 2MHz, specifications apply above 10% of range.
Add 0.001% of reading × (VIN/100V) 2 additional uncertainty for inputs above 100V rms.
Typical values
2001 OHMSTWO-WIRE AND FOUR-WIRE OHMS (2W and 4W Ohms Functions)
Range
FullScale
Resolution
CurrentSource
Resistance Accuracy ±(ppm of reading + ppm of range)
Transfer
24Hours
90Days
1Year
2Years
20 Ω
21.000000Ω
100 nOΩ
7.2 mA
2.5+3
5+4.5
15+6
17+6
20+6
200 Ω
210.00000Ω
1 µΩ
960 µΩ
2.5+2
5+3
15+4
17+4
20+4
2 kΩ
2100.0000kΩ
10 µΩ
960 µΩ
1.3+0.2
2.5+0.3
7+0.4
9+0.4
11+0.4
20 kΩ
21.000000kΩ
100µΩ
96µA
1.3+0.2
2.5+0.3
7+0.4
9+0.4
11+0.4
200 kΩ
210.00000kΩ
1 mΩ
9.6 µA
2.5+0.4
5.5+0.5
29+0.8
35+0.9
40+1
2 MΩ
2.1000000MΩ
10 mΩ
1.9 µA
5+0.2
12+0.3
53+0.5
65+0.5
75+0.5
20 MΩ
21.000000MΩ
100 mΩ
1.4 µA
15+0.1
50+0.2
175+0.6
250+0.6
300+0.6
200 MΩ
210.00000MΩ
1 Ω
1.4 µA
50+0.5
150+1
500+3
550+3
600+3
1 GΩ
1.0500000GΩ
10 Ω
1.4 µA
250+2.5
750+5
2000+15
2050+15
2100+15
OHMS NOTES
Current source has an absolute accuracy of ±5%
For 2-wire mode
Specifications are for 10 power line cycles, 10-reading repeat digital filter, synchronous autozero, autorange off, 4-wire mode, offset compensation on (for 20Ω to 20kΩ ranges), except as noted
For TCAL ±1°C, following 4-hour warm-up. TCAL is ambient temperature at calibration (23°C at the factory).
For TCAL ±5°C, following 4-hour warm-up
Current source is paralleled with a 10MΩ resistance
Specifications apply for 20-reading repeat digital filter, TREF ±0.5°C (TREF is the initial ambient temperature), and for measurements within 10% of the initial measurement value and within 10 minutes of the initial measurement time
2002 DC AMPSDCI INPUT CHARACTERISTICS AND ACCURACY
Range
FullScale
Resolution
MaximumBurdenVoltage
Accuracy ±(ppm of reading + ppm of range)
24Hours
90Days
1Year
2Years
200 µA
210.00000µA
10 pA
0.25 V
50+6
275+25
350+25
500+25
2 mA
2.1000000mA
100 pA
0.3 V
50+5
275+20
350+20
500+20
20 mA
21.000000mA
1 nA
0.35 V
50+5
275+20
350+20
500+20
200 mA
210.00000mA
10 nA
0.35 V
75+5
300+20
375+20
525+20
2 A
2.1000000A
100 nA
1.1 V
350+5
600+20
750+20
1000+20
DC AMPS NOTES
For TCAL ±1°C, following 55-minute warm-up. TCAL is ambient temperature at calibration (23°C at the factory)
For TCAL ±5°C, following 55-minute warm-up
Actual maximum burden voltage = (maximum burden voltage) × (I MEASURED/I FULL SCALE)
2002 AC AMPS
Range
ACI ACCURACY90 Days, 1 Year or 2 Years, TCAL ±5°C, for 5% to 100% of range, ±(% of reading + % of range)
20Hz–50Hz
50Hz–200Hz
200Hz–1kHz
1kHz–10kHz
10kHz–30kHz
30kHz–50kHz
50kHz–100kHz
200 µA
0.35+0.015
0.2+0.015
0.4+0.015
0.5+0.015
2 mA
0.3+0.015
0.15+0.015
0.12+0.015
0.12+0.015
0.25+0.015
0.3+0.015
0.5+0.015
20 mA
0.3+0.015
0.15+0.015
0.12+0.015
0.12+0.015
0.25+0.015
0.3+0.015
0.5+0.015
200 mA
0.3+0.015
0.15+0.015
0.12+0.015
0.15+0.015
0.5+0.015
1+0.015
3+0.015
2 A
0.35+0.015
0.2+0.015
0.3+0.015
0.45+0.015
1.5+0.015
4+0.015
AC AMPS NOTES
Specifications apply for sinewave input, AC+DC coupling, 1 power line cycle, autozero on, digital filter off, following 55-minute warm-up
Add 0.005% of range uncertainty for current above 0.5A rms for self-heating
Typical values
FREQUENCY COUNTER
AC VOLTAGE INPUT
1Hz–15MHz
ACCURACY
±(0.03% of reading)
DC IN-CIRCUIT CURRENT
TYPICAL RANGES
Current
100µA to 12A
Trace Resistance
1mΩ to 10Ω typical
ACCURACY
±(5% + 500µA). For 1 power line cycle, autozero on, 10-reading digital filter, TCAL ±5°C, 90 days, 1 year or 2 years
TEMPERATURE
Built-in linearization for J, K, N, T, E, R, S, B thermocouple types to ITS-90 and 100O platinum RTDs DIN 43760, IPTS-68, and ITS-90.
Keithley 2001/MEM1 Product Specifications Summary:(for full specs please see the full specs tab or the data sheet)